CONTINUOUS PROCESS ANALYZER FOR SULFUR DIOXIDE BY INFRARED DETECTOR, Model : 460 SERIES, Brand : Nova
Keterangan
The Nova 460 Series SO2 analyzers utilize a reliable microprocessor-based non-dispersive infrared detector specific to SO2. In operation, the internal sample pump draws the sample gas through the filter/ condensate trap, secondary filter, flow meter, and then on to the infrared detector. The detector output is digitally linearized and then displayed as percent or PPM SO2 on the digital meter. The output is also available as an analog or digital signal.
Specifications
Method of Detection
Microprocessor based single cell NDIR
Ranges Available
460 - 0-1, 0-5, 0-10, 0-20, 0-30, 0-50, 0-100% SO2
460L - 0-1000, 0-3000, 0-5000, PPM SO2
Resolution 0.1% SO2 on % versions, 10 PPM SO2 on PPM versions
Accuracy & Repeatability 1% of full scale
Drift Less than 2% of full scale per month
Response Time ( T-90)
Less than 15 seconds
Ambient Temperature Range
40 to 122° F ( 5 to 50° C) , Lower temperatures ( -5° F, -20° C) with Cold Weather Package
Linearity ± 1% of full scale
Size & Weight Physical data will vary depending on enclosure style and options required
Power 115 VAC, 60 Hz - 220 VAC, 50 Hz
Output Options 4-20 mA into 500 ohms non-isolated standard
Isolated 4-20 mA, RS232, RS485, MODBUS, Ethernet outputs
Alarms High and/ or low alarm contacts available, relay contacts SPDT 5A @ 220 VAC rating, Low flow alarm
Analyzer Type Continuous Online, Single Gas
Hazardous Area Rating Options
General Purpose, Non-Hazardous
Applications
For continuous analysis of sulfur dioxide ( SO2) in process gas streams using an infrared detector. Percent and PPM levels available as Models 460 & 460L respectively.
Features
Percent ( 460) or PPM ( 460L) SO2 by infrared detector
Detector is sensitive, reliable, and non-depleting
Bright digital readout
Fast response ( T90 - less than 15 seconds)
Microprocessor control for push button calibration
Easy-to-maintain modular layout
Built in sample pump, flowmeter, and filters
4-20 mA outputs
Options
Hi/ Low gas, low flow, and diagnostic alarms available
Isolated analog, RS232, RS485, MODBUS, and Ethernet outputs available
Sample conditioning systems available for removal of moisture, acidic gas, and dust
Cabinet purge system available for use in hazardous areas
One-step ' cal-now' button with touch screen LCD display
Full automatic calibration with touch screen LCD display
Cold weather package for operation to -5° F ( -20° C)
Cabinet coolers can be fitted to most models
Sensors mounted in temperature controlled heated box to improve accuracy
High temperature filters and probes
Calibration
Easily zero checked on ambient air
Span using a known SO2 in nitrogen ( N2) calibration gas.